· Source Emissions Monitoring - Abatement System Efficiency
· Many other applications, contact us to discuss your application
· Detector Flame Ionization Detector (FID)
Oxidizer Oxidizes non-methane hydrocarbons Ranges Analog outputs are user-definable for any full scale value (based upon calibration) within these overall ranges: MDQ 0.02 ppm, full-scale 50 ppm (Air balance) - MDQ is typically <20 ppb as CH4 with signal to noise ratio of 3:1 MDQ 0.06 ppm, full-scale 500 ppm (Air balance) MDQ 0.150 ppm, full-scale 5000 ppm (O2 balance) Repeatability +/- 1% full-scale response Drift +/- 1% of full-scale over 24 hours Response Time <30 seconds to 90% of final reading Alarms Multilevel concentration and fault alarms that result in audible and visual alarms Alarms may also be mapped to relays to control external equipment Sampling Internal single module for pre-filtered (<0.1 μ) non-condensing samples, with or without sample pump Calibration Programmable automatic or manual calibration Support Gas Hydrogen 35 cc/min. (H2/He blend 100 cc/min.), Air 175 cc/min. (typical) Hydrocarbon content must be less than 1 ppm Fuel blend options available, consult MOCON-Baseline Power 90 – 230 VAC, 50/60 Hz, 3A Relay Outputs 5 programmable form A relays rated to 3 A @ 230 V AC (optional additional 9 relay output board) Analog Outputs 1 programmable 0-20 mA or 4-20 mA isolated output (optional additional 3 analog output board) Digital Outputs RS-232, Ethernet